JavaScript is disabled for your browser. Some features of this site may not work without it.

Validita přístroje VO2 Master Pro pro predikci maximální spotřeby kyslíku a měření klidového metabolismu

Validity of the VO2 Master Pro device for maximal oxygen consumption prediction and the measurement of resting metabolism

Thumbnail
View/Open
review_142056.html (9.019Kb)
final-thesis.pdf (4.854Mb)
appendix-1.zip (2.312Mb)
Posudek-Vedouci prace-Posudek Goldschmidt_bakalarka_2022.pdf (92.65Kb)
Author
Goldschmidt, Tobiáš
Advisor
Chlíbková, Daniela
Referee
Nohel, Michal
Grade
A
Altmetrics
Metadata
Show full item record
Abstract
Meranie maximálnej spotreby kyslíku a kľudového metabolizmu sú jedným zo základných stavebných kameňov v testovaní športovcov a medicínskych pacientov. Na toto meranie sa využívajú analyzátory dýchacích plynov. Tieto analyzátory sú väčšinou neprenosné nálkadné zariadenia. VO2 Master Pro je na druhej strane prenosný prístroj s rádovo nižšou cenou. Jeho nevýhodou je však že obshuje len O2 senzor a nie aj CO2 senzor narozdiel od prístrojov chápané ako medicínsky štandard. Cieľom tejto práce bolo stanoviť validitu a odchýlku prístroja VO2 Master Pro pre meranie maximálnej spotreby kyslíku a kľudového metabolizmu. Odchýlka pre meranie maximálnej spotreby kyslíku bola 4,15 ± 4,41 %. Pre meranie kľudového metabolizmu bola odchýlka 41,20 ± 11,44 %. Pre meranie maximálnej spotreby kyslíku je VO2 Master Pro validný ale pre meranie kľudového metabolizmu validný nie je. Takisto bolo dokázané že veľkosť nameranej hodnoty má nelieneárny vzťah ku veľkosti odchýlky pri meraní maximálnej spotreby kyslíku.
 
Measurements of maximal oxygen uptake and resting metabolic rate are one of the most important things in testing athletes and madical patients. For these measurements are used breath analyzers. These analyzers are usually non-portable costly devices. VO2 Master Pro is on the other hand portable breath analyzer with much lower prize. It is biggest disadvantage is that it contains only O2 sensor and not CO2 sensor as madically aproved devices. Aim of this study is to determinate validity and difference of measurements between VO2 Master Pro and reference device for measuring maximal oxygen uptake and resting metabolic rate. Difference for measuring maximal oxygen uptake was 4,15 ± 4,41 %. For measuring of resting metabolic rate difference was 41,20 ± 11,44 %. For measuring of maximal oxygen uptake is VO2 Master Pro valid but for measuring of resting metabolic rate it is not. It was also confirmed, that size of measured value have non-linear corelation with size of difference in measuring maximal oxygen uptake.
 
Keywords
VO2 Master Pro, maximálna spotreba kyslíku, kľudový metabolizmus, VO2peak, VO2max, RMR, validita, VO2 Master Pro, maximal oxygen uptake, resting metabolic rate, VO2peak, VO2max, RMR, validity
Language
čeština (Czech)
Study brunch
bez specializace
Composition of Committee
doc. Ing. Jana Kolářová, Ph.D. (předseda) Ing. Lukáš Smital, Ph.D. (místopředseda) Mgr. Daniela Chlíbková, Ph.D. (člen) Mgr. Ivan Struhár, Ph.D. (člen) Ing. Oto Janoušek, Ph.D. (člen)
Date of defence
2022-06-20
Process of defence
Student prezentoval výsledky své práce a komise byla seznámena s posudky. Ing. Janoušek položil otázku: Jak jste vybíral použité statistické testy? Nešlo by použít silnější statistické testy? Byl jste schopný proměřit celý rozsah použitých přístrojů? Uvádí výrobce rozsah měřených hodnot? Mgr. Struhár položil: Jaký stav považujete za plató a jaká je jeho hodnota? Student obhájil bakalářskou práci a odpověděl na otázky členů komise a oponenta.
Result of the defence
práce byla úspěšně obhájena
Persistent identifier
http://hdl.handle.net/11012/207471
Source
GOLDSCHMIDT, T. Validita přístroje VO2 Master Pro pro predikci maximální spotřeby kyslíku a měření klidového metabolismu [online]. Brno: Vysoké učení technické v Brně. CESA. 2022.
Collections
  • 2022 [16]
Citace PRO

Portal of libraries | Central library on Facebook
DSpace software copyright © 2002-2015  DuraSpace
Contact Us | Send Feedback | Theme by @mire NV
 

 