Calibrating the Six-Port Reflectometer Using a Matched Load And Four Unity-Reflection Standards (Part 2)
Abstrakt
A convenient and exact method of using a sliding load to the calibration of the single six-port reflectometer is described. Neither reflection coefficient nor positions of the sliding load settings need be known. Reflection coefficient of the load is not limited to small values. No numerical iterations are involved in the calibration.
Klíčová slova
Six-port reflectometer, calibrationTrvalý odkaz
http://hdl.handle.net/11012/58634Typ dokumentu
Recenzovaný dokumentVerze dokumentu
Finální verze PDFZdrojový dokument
Radioengineering. 1992, vol. 1, č. 2, s. 2-8. ISSN 1210-2512http://www.radioeng.cz/fulltexts/1992/92_02_01.pdf