Mechanická analýza vlivu výrobních odchylek na styk hlavice a jamky totální endoprotézy kyčelního kloubu

Mechanical Analysis of the Influence of Fabrication Tolerances on Contact Conditions between the Acetabular Cup and the Femoral Head in Total Hip Replacement

Thumbnail
View/Open
final-thesis.pdf (3.921Mb)
thesis-1.pdf (4.576Mb)
review_34081.html (1.946Kb)
Author
Koukal, Milan
Advisor
Florian, Zdeněk
Referee
Horyl, Petr
Repko, Martin
Janíček, Přemysl
Grade
P
Altmetrics
Metadata
Show full item record
Abstract
V oblasti endoprotetické chirurgie se vyskytuje nezanedbatelné procento implantátu vykazujících vady, které vedou k selhání celé protézy. Jedním z typů selhání totální kyčelní endoprotézy je uvolnění jamky z pánevní kosti. Tato studie posuzuje vliv výrobních odchylek jako jeden z možných vlivů vedoucích k uvedenému typu selhání protézy. Posuzovány byly rozměrové a geometrické odchylky kulových ploch keramické hlavice a polyethylenové jamky. Bylo zjištěno, že tyto odchylky mají významný vliv na stykové poměry mezi hlavicí a jamkou. Posuzované veličiny byly kontaktní tlak a třecí moment, což jsou veličiny ovlivňující životnost celé protézy. Pro nalezení řešení byla použito výpočtového modelování. Z výsledků analýzy řešení lze doporučit předepisovat výrobní tolerance tak, aby mezi hlavicí a jamkou vzniklo uložení s vůlí, kdy interval vhodných hodnot vůle je 0 mm až 0,05 mm. Lze důrazně nedoporučit uložení s přesahem z důvodu výrazně zhoršených stykových podmínek. Odchylka kruhovitosti hlavice by neměla být vyšší než 0,025 mm.
 
In endoprosthesis surgery there are typically a high percentage of implant defects, these can lead to failure of the whole prosthesis. One type of total hip replacement function loss is acetabular cup loosening from the pelvic bone. This disertation examines manufacture perturbations as one of the possible reasons for this kind of failure. Both dimension and geometry manufacturing perturbations of ceramic head and polyethylen cup were analyzed. We find that perturbations in the variables analysed here affect considered values of contact pressure and frictional moment. Furthermore, contact pressure and frictional moment are quantities affecting replacement success and durability. From obtained results it can be recommended to fit head and cup with a clearance of between 0 mm and 0.05 mm. It can not be recommend using interference type of fit because of strong deterioration of the contact conditions. Roundness perturbation of ceramic head should not exceed 0.025 mm.
 
Keywords
Totální kyčelní endoprotéza, MKP, kruhovitost, kontaktní tlak, třecí moment, biomechanika, Total hip replacement, FEM, roundness, contact pressure, frictional moment, biomechanics
Language
čeština (Czech)
Study brunch
Inženýrská mechanika
Composition of Committee
prof. Ing. Jindřich Petruška, CSc. (předseda) prof. Ing. Petr Horyl, CSc. (člen) MUDr. Martin Repko, Ph.D. (člen) prof. Ing. Přemysl Janíček, DrSc. (člen) doc. Ing. Jana Horníková, Ph.D. (člen) doc. Ing. Vladimír Fuis, Ph.D. (člen)
Date of defence
2010-12-20
Process of defence
Zkušební komise konstatuje, že předložená DP řeší vysoce aktuální téma současného biomechanického výzkumu a její závěry jsou přínosem pro klinickou praxi. Doktorand v plném rozsahu splnil vytčené cíle a jeho závěry prohlubují poznání v oblasti konstrukce i aplikace totálních endoprotéz kyčelního kloubu.
Result of the defence
práce byla úspěšně obhájena
Persistent identifier
http://hdl.handle.net/11012/4376
Source
KOUKAL, M. Mechanická analýza vlivu výrobních odchylek na styk hlavice a jamky totální endoprotézy kyčelního kloubu [online]. Brno: Vysoké učení technické v Brně. Fakulta strojního inženýrství. 2010.
Collections
  • 2010 [55]
Citace PRO

Portal of libraries | Central library on Facebook
DSpace software copyright © 2002-2015  DuraSpace
Contact Us | Send Feedback | Theme by @mire NV
 

 

Browse

All of repositoryCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsThis CollectionBy Issue DateAuthorsTitlesSubjects

My Account

LoginRegister

Statistics

View Usage Statistics

Portal of libraries | Central library on Facebook
DSpace software copyright © 2002-2015  DuraSpace
Contact Us | Send Feedback | Theme by @mire NV