Mathematical model of Mobile Circulatory Module for ex-vivo lungs perfusion

dc.contributor.authorŠolc, Františekcs
dc.contributor.authorZezulka, Františekcs
dc.contributor.authorVeselý, Ivocs
dc.coverage.issue4cs
dc.coverage.volume48cs
dc.date.accessioned2020-08-04T11:00:07Z
dc.date.available2020-08-04T11:00:07Z
dc.date.issued2015-05-13cs
dc.description.abstractThe article deals with mathematical modelling of a system for ex-vivo lung perfusion. The system which is modelled here is named Mobile Circulatory Module and is designed for transport of live lungs namely porcine or human lungs. Transported lungs should be used for various research or even for transplantation. The article describes mathematical model of thermal circuit and hydraulic circuit which are vital parts of the MCM to secure living conditions of the transported organ. Mathematical model is simulated with MATLAB Simulink and shows good qualitative correspondence with real behavior of the system. Based on math model there is designed control algorithm for temperature control and there is discussed control of perfusion liquid as well.en
dc.description.abstractPříspěvek se zabývá matematickým modelováním systému pro prokrvení plic ex-vivo. Tento systém, který je zde modelován je pojmenována Mobile Circulatory Module a je určen pro přepravu živých plic zejména prasat nebo lidské plíce. Přepravované plíce by měly být použity pro různé výzkumné učely, nebo dokonce k transplantaci.Článek popisuje matematický model tepelného okruhu a hydraulickém okruhu, které jsou životně důležité části MCM k zajištění životních podmínek dopravovaného orgánu. Matematický model je simulován s MATLAB Simulink a vykazuje dobrou kvalitativní korespondenci s skutečného chování systému. Z matematickém modelu lze usuzovat, že je určen řídicí algoritmus pro regulaci teploty, a je zde diskutována kontrola perfusním kapaliny.cs
dc.formattextcs
dc.format.extent392-397cs
dc.format.mimetypeapplication/pdfcs
dc.identifier.citationIFAC-PapersOnLine (ELSEVIER). 2015, vol. 48, issue 4, p. 392-397.en
dc.identifier.doi10.1016/j.ifacol.2015.07.066cs
dc.identifier.issn2405-8963cs
dc.identifier.other115453cs
dc.identifier.urihttp://hdl.handle.net/11012/186777
dc.language.isoencs
dc.publisherIFACcs
dc.relation.ispartofIFAC-PapersOnLine (ELSEVIER)cs
dc.relation.urihttps://www.sciencedirect.com/science/article/pii/S2405896315008411cs
dc.rights(C) IFACcs
dc.rights.accessopenAccesscs
dc.rights.sherpahttp://www.sherpa.ac.uk/romeo/issn/2405-8963/cs
dc.subjectplíce
dc.subjectorgany
dc.subjectMCM
dc.subjectmodel v Matlabu
dc.subjectřízení průtoku
dc.subjectřízení teploty
dc.subjectex-vivo lungs perfusionen
dc.subjectisolated organen
dc.subjectMCMen
dc.subjectMatlab modelen
dc.subjectcontrol of flowen
dc.subjecttemperature controlen
dc.titleMathematical model of Mobile Circulatory Module for ex-vivo lungs perfusionen
dc.title.alternativeMatematický model mobilního výživového modulu pro plícecs
dc.type.driverarticleen
dc.type.statusPeer-revieweden
dc.type.versionpublishedVersionen
sync.item.dbidVAV-115453en
sync.item.dbtypeVAVen
sync.item.insts2020.08.04 13:00:07en
sync.item.modts2020.08.04 12:28:51en
thesis.grantorVysoké učení technické v Brně. Fakulta elektrotechniky a komunikačních technologií. Ústav automatizace a měřicí technikycs
thesis.grantorVysoké učení technické v Brně. Fakulta elektrotechniky a komunikačních technologií. Centrum výzkumu a využití obnovitelných zdrojů energiecs
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
1s2.0S2405896315008411main.pdf
Size:
641.82 KB
Format:
Adobe Portable Document Format
Description:
1s2.0S2405896315008411main.pdf